| Scan Mode | Helical |
| Collimation | 0.5mm x 40 |
| kVp | 120 |
| mAs | 240 |
| Rotation Time | 0.75s |
| Scan Range | 140.0mm |
| Reconstruction | AiCE Brain LCD |
| CTDIvol | 57.0mGy |
| DLP | 915.8mGy·cm |
| Effective Dose | 5.4mSv |
| k-factor | 0.0021 (AAPM Report 96) |
Auto subtraction demonstrates intracranial aneurysms associated with AVM.
View Scan Parameters (Non-Contrast)| Scan Mode | Ultra-Helical |
| Collimation | 0.5mm x 80 |
| kVp | 120 |
| mAs | 165 |
| Rotation Time | 0.75s |
| Scan Range | 225.0mm |
| Reconstruction | AiCE |
| CTDIvol | 51.5mGy |
| DLP | 953.1mGy·cm |
| Effective Dose | 2.00mSv |
| k-factor | 0.0021 (AAPM Report 96) |
| Scan Mode | Ultra-Helical |
| Collimation | 0.5mm x 80 |
| kVp | 120 |
| mAs | 165 |
| Rotation Time | 0.75s |
| Scan Range | 225.0mm |
| Reconstruction | AiCE |
| CTDIvol | 35.2mGy |
| DLP | 698.8mGy·cm |
| Effective Dose | 1.46mSv |
| k-factor | 0.0021 (AAPM Report 96) |
CTA with SURESubtraction demonstrates a right ICA occlusion.
View Scan Parameters (Non-Contrast)| Scan Mode | Ultra-Helical |
| Collimation | 0.5mm x 80 |
| kVp | 120 |
| mAs | 150 |
| Rotation Time | 0.75s |
| Scan Range | 164.0mm |
| Reconstruction | AiCE |
| CTDIvol | 42.1mGy |
| DLP | 871.4mGy·cm |
| Effective Dose | 1.83mSv |
| k-factor | 0.0021 (AAPM Report 96) |
| Scan Mode | Ultra-Helical |
| Collimation | 0.5mm x 80 |
| kVp | 120 |
| mAs | 187 |
| Rotation Time | 0.75s |
| Scan Range | 164.0mm |
| Reconstruction | AiCE |
| CTDIvol | 52.7mGy |
| DLP | 1089.3mGy·cm |
| Effective Dose | 2.28mSv |
| k-factor | 0.0021 (AAPM Report 96) |
AiCE DLR enabling high diagnostic confidence in evaluation of small bony structures in IAC, through significant noise reduction while maintaining high-resolution.
View Scan Parameters| Scan Mode | Ultra-Helical |
| Collimation | 0.5mm x 40 |
| kVp | 120 |
| mAs | 150 |
| Rotation Time | 0.5s |
| Scan Range | 57.0mm |
| Reconstruction | AiCE Inner Ear |
| CTDIvol | 35.6mGy |
| DLP | 181.3mGy·cm |
| Effective Dose | 0.4mSv |
| k-factor | 0.0021 (AAPM Report 96) |
SURESubtraction automatically generates images with subtracted bones improving the visualization of the neurovascular anatomy.
View Scan Parameters (Non-Contrast)| Scan Mode | Ultra-Helical |
| Collimation | 0.5mm x 80 |
| kVp | 100 |
| mAs | SUREExposure |
| Rotation Time | 0.5s |
| Scan Range | 350.0 mm |
| Reconstruction | AiCE |
| CTDIvol | 3.9mGy |
| DLP | 152.7mGy·cm |
| Effective Dose | 0.47mSv |
| k-factor | 0.0031 (AAPM Report 96) |
| Scan Mode | Ultra-Helical |
| Collimation | 0.5mm x 80 |
| kVp | 100 |
| mAs | SUREExposure |
| Rotation Time | 0.5s |
| Scan Range | 350.0 mm |
| Reconstruction | AiCE |
| CTDIvol | 3.9mGy |
| DLP | 152.7mGy·cm |
| Effective Dose | 0.80mSv |
| k-factor | 0.0031 (AAPM Report 96) |
Contrast enhanced CT of the head with AIDR 3D demonstrates a brain hemorrhage.
View Scan Parameters| Scan Mode | Ultra-Helical |
| Collimation | 0.5mm x 80 |
| kVp | 120 |
| mAs | 225 |
| Rotation Time | 0.5s |
| Scan Range | 162.0 mm |
| Reconstruction | AIDR 3D |
| CTDlvol | 27.2mGy |
| DLP | 561.0mGy·cm |
| Effective Dose | 1.17mSv |
| k-factor | 0.0021 (AAPM Report 96) |
Iodine maps,110 Equiv. kV Enhanced images demonstrate tumor enhancement, the VNC images show bone destruction.
View Scan Parameters| Scan Mode | DE-Helical |
| Collimation | 0.5mm x 80 |
| kVp | 80/135 |
| mAs | 100/570 |
| Rotation Time | 0.5s |
| Scan Range | 310.0mm |
| Reconstruction | AIDR 3D |
| CTDIvol | 21.4mGy |
| DLP | 688.8mGy·cm |
| Effective Dose | 1.14mSv |
| k-factor | 0.0021 (AAPM Report 96) |