Average Value | Pre Hi-Def Era | Post Hi-Def Era | % Diff | p-value |
---|---|---|---|---|
Procedure time (min) | 43.15 | 39.18 | -9.2% | < 0.001 |
Cumulative Air Kerma (Gy) | 1.11 | 1.00 | -9.6% | 0.003 |
Fluoro Time (min) | 19.12 | 17.48 | -8.6% | 0.004 |
# of Pulses | 17,643 | 16,192 | -8.2% | 0.005 |
# of 3D Spins | 0.30 | 0.26 | -12% | 0.023 |
# of Irradiation events | 85.0 | 82.2 | -3.3% | 0.059 |
References
1. Setlur Nagesh SV, et al., Proc. SPIE 10573, Medical Imaging 2018: Physics of Medical Imaging, 1057365 (9 March 2018).
2. Setlur Nagesh SV, et al., J Neuroimaging. 2019 Sep;29(5):565-572.
3. Setlur Nagesh SV, et al., AJNR Am J Neuroradiol. 2019 Feb;40(2):302-308.
4. Setlur Nagesh SV, et al., J Neurointerv Surg. 2020 Sep;12(9):897-901.
5. Krebs J, et al., Proc. SPIE 10948, Medical Imaging 2019: Physics of Medical Imaging, 109480E (1 March 2019).
6. Miller DL, et al., J Vasc Interv Radiol. 2003 Jun;14(6):711-27.